This instrument is widely supplied in measuring axunge paint plastic, food, medication, adhesive, cosmetic and so on like fluidic sticky resistance and liquid absolute viscosity.fix and measuring must be done according requirement from instruction book。
 
Main technic spec
 
·Range of measure:10~105mpa.s
·Rotate speed: 6,12,30,60turn/minute
·Error of measure:±5%(Newton liquid)
·Power:220V±10%
         50Hz
 
Fix

1.     take out chassis base
andsupport from packing
boxcircumrotate three
level 。
2.adjust foot into tapping
hole in chassis base(chart 1)
3.main machine fix:
a. take out main machine by unscrewing the screws in chassis
b. unscrew the handle root screw . inset the power line into handle slot(chart2)
c. penetrate the handle into hole under slot(chart2)
d. screw the handle fix knob to screw the handle and keep the main machine level mostly (chart3)
 
 
8.measure the temperature of measured liquid directly , do not use the temperature value of constant temperature bath
9. it is normal when the viscosity changes because of change of speed and length of measure time during measuring. That is not the instrument hitch.
10.operate seriously according the operation manual
 
 

The whole set

1.main machine————————1
2.(1,2,3,4)rotors———— 1
3.protective support—————— 1
4.support——————————  1
5.lever adjustment screw———— 3
6.manual——————————  1
7.certificate ————————— 1
 
Quotiety table
 
     Run/minute
rotor
60 30 12 6
1 1 2 5 10
2 5 10 25 50
3 20 40 100 200
4 100 200 500 1000
 

 

Notice

1.this instrument should be used in 10—30
2.protective support must be fixed when measuring, or the viscosity value will change as the change of diameter of container
3.the rotor connective screw staff should be raised when fix or remove rotor, or the head of shaft will be damaged. Keep the bleb from rotor when rotor is immerged into liquid
4.the instrument should not be placed in oblique or inverse direction. Especially the rotor is fixed.
5.after fixing the rotor, do not turn on machine to turn the rotor without immerging the rotor into liquid
6.remove rotor carefully after finishing work every time.  Clear it extremely, dry is with blower, and put it back on the rotor support in vertically
7.pointer lock staff should be packaged by yellow package cover and rubber hand when carrying the instrument.
 


 
e. remove the rubber hand from pointer lock staff and yellow package cover from main machine inferior part.
alarm: do not turn on machine before removing the above two objects or it will damage the machine
power on, adjust the level adjustment foot(chart4). Make the bleb in the middle of level set(chart5).
 
Operation
 
1.first rightly rotate the protective support into the instrument inferior part.
2.prepare the liquid for measuring, select the column container having diameter not less than 70 mm
3.control the temperature in liquid for measuring correctly according the precision required by measuring
4. select the right rotor number by estimating the approximate range of viscosity of liquid or measuring, and rotate left the rotor into the connective screw staff(chart4),then rotate the up and down knob(chart5),make the instrument down slowly, immerging into measured liquid gradually, until the rotor liquid level balances the liquid level(chart4).put down the pointer lock staff, power on, relax.

the pointer lock staff,
to make the rotor
circumrotate in the
liquid. The speed is
20-30 run, until the
pointer stay at one
position . put down
the pointer lock staff to
make the reading unchangeable,
then turn off the motor to
make the pointer stay
in the range of reading
window,then get the
reading directly. If the
pointer do not stay in
the frame after turning
off motor, go on to put
down the pointer lock
staff. Repeat the action 
of turning on and off to
make the pointer in the
frame of window. If
selecting 6 and 12 turn /minute,
get the reading directly
without locking the pointer.5.the best is that the read pointer stay 30—90 division range, or change the rotor number or run speed to get the requirement 6.when you can not predict the viscosity value of liquid for measuring , the rotor and turn speed knob should be selected by turning from little value to big value, and from slow to fast, to measure the viscosity value.
 
 
Quotiety  and error correcting
1.the absolute viscosity value of liquid measured can be get by multiplying the reading pointed on dial when measuring and quotiety in quotiety table
that is:
η(mpa.s=k×α
k= quotiety
α=pointer reading
η=viscosity value
2.frequency error correcting
                                     normal frequency
actual viscosity value= indicated value ×—————————
                              actual frequency
 
 

 

Range stable and quotiety table

range stable
rotor Turn/minute
60 30 12 6
1 100 200 500 1000
2 500 1000 2500 5000
3 2000 4000 10000 20000
4 10000 20000 50000 100000
 
 
 

 


[上一篇:烏氏粘度計(jì)原理及其應(yīng)用 ]    [下一篇:美國(guó) BROOKFIELD博勒飛粘度計(jì)使用說(shuō)明書 ]

使用方法

常見問(wèn)題

瀏覽更多同類產(chǎn)品

STM-IVB斯托默粘度計(jì)
STM-IVB斯托默粘度計(jì)
 NDJ-1E 旋轉(zhuǎn)粘度計(jì)
NDJ-1E 旋轉(zhuǎn)粘度計(jì)
NDJ-1D 布氏旋轉(zhuǎn)粘度計(jì)
NDJ-1D 布氏旋轉(zhuǎn)粘度計(jì)
烏氏粘度計(jì)
烏氏粘度計(jì)
国产呻吟久久久久久久92| 色天使色偷偷色噜噜| 日本xxxx色视频在线观看免费| 免费a片看黄网站www| 日韩在线观看精品亚洲| 国产97人人超碰cao蜜芽prom | 无码办公室丝袜ol中文字幕| 亚洲电影天堂在线国语对白| 收集最新中文国产中文字幕| 中文字幕第55页一区| 亚洲国产精品第一区二区| 午夜精品久久久久久99热| 午夜亚洲av永久无码精品| 99精品人妻无码专区在线视频区| 一个人看的www免费视频中文| 欧美孕妇变态重口另类| 国产亚洲精品精品精品| 色噜噜狠狠综曰曰曰| 国产又粗又猛又爽又黄91精品| 久久久无码精品亚洲日韩蜜臀浪潮 | 亚洲人成色99999在线观看| 国产精品99久久久精品无码| 女人与公拘交酡全过程| 亚洲av美国av产亚洲av图片| 欧美人牲交免费观看| 亚洲多毛妓女毛茸茸的| 一本丁香综合久久久久不卡网站 | 成人午夜福利视频| 久久免费看黄a级毛片| 日韩精品久久久久久免费| 日韩精品无码免费专区网站| 国产一区二区三区在线观看免费| 青青国产线免观| 国精产品一二三区精华液| 色诱视频在线观看| 最近2019中文字幕大全第二页| 国产精品∧v在线观看| 国内老熟妇乱子伦视频| 人妻一本久道久久综合久久鬼色| 国产成人精品视频网站| 精品久久无码中文字幕|